Online Support for the RapidMatch® XI Spectrophotometer
All the information you need to set up, use and troubleshoot your RAPIDMATCH® XI device.

All the information you need to set up, use and troubleshoot your RAPIDMATCH® XI device.
Prior to contacting X-Rite support department for instrument problems, try the applicable solution(s) described below. If the condition persists, contact us using one of the methods listed in the Service Information section.
| Cause | Solution | 
| Instrument is in power down mode. | Press the power button. | 
| Battery pack is very low or bad. | Charge the battery. If battery pack is bad, replace the battery pack.  | 
| No battery pack installed. | Install batteries or plug in power adapter. | 
| Reset the instrument from the Settings>Systems & Diagnostics>Systems Information screen. | 
| Cause | Solution | 
| Material being measured is damaged (e.g. scratched) | Obtain new material. | 
| Instrument requires calibration. | Refer to the Calibration procedure. |